CLINICAL STUDIES |
Department of Paediatrics, Institute of Clinical Sciences, The Sahlgrenska Academy at Göteborg University, S-416 85, Göteborg, Sweden
(Correspondence should be addressed to C Ankarberg-Lindgren;; Email: carina.ankarberg-lindgren{at}vgregion.se)
| Abstract |
|---|
|
|
|---|
Methods: The lower limit of detection and clinical usefulness (functional sensitivity) of three commercial estradiol immunoassays were validated by use of 500 sera from prepubertal and pubertal children and 55 pooled sera. The three immunoassays consisted of two modified direct immunoassays; one RIA (Spectria Estradiol RIA) and one time-resolved fluoroimmunoassay (AutoDELFIA Estradiol), both with increased serum volume in relation to antibody concentration and extended incubation time. In the third method, serum was purified and concentrated using diethyl ether extraction prior to measurement by the modified Spectria Estradiol RIA.
Results: The lower limits of detection and clinical usefulness were 9 and 30 pmol/l for the direct RIA, 11 and 50 pmol/l for the AutoDELFIA, and 4 and 6 pmol/l for serum determined by extraction RIA. When measuring the serum pool originating from girls at breast stages 1–2, the direct RIA and AutoDELFIA resulted in significantly higher 17β-estradiol concentrations when compared with the extraction RIA (+58 and +267%, P<0.001). We found a significant difference in 17β-estradiol concentrations between girls at breast stages 1 (median 6 pmol/l) and 2 (median 16 pmol/l), when quantified by the extraction RIA (P<0.0001) but no difference when quantified with the direct RIA (median values 12 and 14 pmol/l respectively).
Conclusion: For determination of low serum 17β-estradiol concentrations in children, an extraction step prior to commercial immunoassay is needed to achieve clinically useful results.
| Introduction |
|---|
|
|
|---|
Measurements of 17β-estradiol obtained by mass spectrometry identification are considered as the reference method (1, 2, 3). However, until mass spectrometry is practical for widespread use, the most widely used technique for the measurement of serum 17β-estradiol is and will be immunoassays, due to its simplicity and reliability. A trend is to use high-performance automated instruments with use of small specimen volumes. For the most part, these commercial assays are designed for fertility investigations in female adults, but they are not sensitive enough for quantifying the low levels present in children (4), men, and postmenopausal women (3, 5). Despite improvements regarding sensitivity during recent years, direct assays have not been thoroughly validated for low 17β-estradiol concentrations in serum and have not reached the degree of sensitivity and specificity in clinical practice as expected from the manufacturers' documentation. It has been shown that in samples from postmenopausal women with low 17β-estradiol, extraction RIAs correlate better with mass spectrometry than direct assays do (3).
In this study, we evaluate two sensitive commercial 17β-estradiol immunoassays for clinical usefulness in prepubertal and pubertal children. As our standard for comparison, we use an extraction RIA which we have used in our clinical practice for a decade (6).
| Materials and methods |
|---|
|
|
|---|
All samples used in this study are serum samples submitted for routine clinical testing in children. After 17β-estradiol measurements for routine clinical testing, identification markings were removed from all samples. In the comparison between the direct RIA and extraction RIA, pubertal stages according to Tanner (7) and ages were registered in addition to 17β-estradiol values.
Assay procedure
Spectria direct RIA The original assay was a coated tube RIA (Spectria Estradiol RIA, Orion Diagnostica, Espoo, Finland) with lower detection limit 30 pmol/l (defined as the apparent concentration two S.D. from the counts at maximum binding). The cross-reactivity of the estradiol antiserum had been tested by the manufacture for 30 estradiol-related steroids. The cross-reactivity was below 1% for estrone, 16-oxestradiol, estradiol-3-glucuronide, estriol, and 16-hydroxy estrone, below 0.1% for progesterone, corticosterone, 2-hydroxyestradiol, and below 0.001% for the rest of the related steroids.
To increase the sensitivity of the RIA, 150 µl serum was used instead of 100 µl. The incubation time was extended from 2 h at 37 °C to 16–19 h at room temperature. In addition to the kit calibrators with the lowest concentration of 50 pmol/l, three calibrators of lower concentrations (6.25, 12.5, and 25 pmol/l) were prepared by dilution of the 50 pmol/l calibrator with the zero calibrator. Otherwise, the RIA was conducted according to the manufacturer's instructions.
Spectria extraction RIA In the extraction RIA, serum was purified and concentrated using diethyl ether extraction prior to quantifying the 17β-estradiol concentrations by Spectria. The basic method has been published previously (6, 8), although the latest revision has not been published before. In the present study, 4 ml diethyl ether (high performance liquid chromatography grade, Merck 1.00921) was added by a dispenser (Hirschmann Laborgeräte, Eberstadt, Germany) to 450 µl serum in 10 ml glass tubes with a tight stopper (rubber, coated with Teflon). For confirmation of the detection limit when the aim was to only purify the serum, 2 ml diethyl ether was added to 200 µl serum in 10 ml glass tubes. The sex steroids were extracted to the ether phase for 30 min in a rotation mixer. The serum phase was thereafter frozen by putting the tubes in a beaker with a mixture of 99.5% ethanol and dried ice. The ether phase was transferred to 5 ml glass tubes, and totally evaporated on a 37 °C heat block under a stream of nitrogen. Reconstitution was made using 200 µl zero calibrator (Spectria Estradiol zero calibrator, Orion Diagnostica). After incubation at 37 °C for 20 min and vortexing, the samples were ready for quantification by the modified Spectria Estradiol RIA, using 150 µl. The entire extraction procedure was done in duplicate.
To assess recovery of the extraction procedure, serum samples from ten different children were incubated with a small amount of tritiated 17β-estradiol for 2–3 h in a shaking water bath (37 °C). The extraction procedure was then performed as above. The radioactivity was determined in 150 µl of the final sample volume. The recovery of 17β-estradiol using the extraction procedure was 91±1% (mean±S.D.), which was close to the theoretical recovery of 90%. These results are in line with our earlier estimation of the extraction procedure (6). All results have been corrected for the extraction loss.
AutoDELFIA assay The original assay is an automated time-resolved fluoroimmunoassay (AutoDELFIA Estradiol, Wallac oy, Turku, Finland) with a lower detection limit 50 pmol/l, as stated by the manufacturer and Brito et al. (9). Required sample volume is 25 µl. To increase the sensitivity of the assay, the antibody was diluted 1:18 instead of 1:10 and incubation time was extended to 2 h. In addition to the kit calibrators; with the lowest concentration of 50 pmol/l, one calibrator of lower concentrations (15 pmol/l) was prepared by dilution with a zero calibrator. Otherwise, the AutoDELFIA was conducted according to the manufacturer's instructions.
Assay evaluation
Spectria assay To estimate the lower analytical detection limit, replicates of 20 zero calibrators were measured in three different assays of different kit lots. The analytical sensitivity was confirmed in dilution tests in which 100 serum samples from children were purified by the extraction procedure using 200 µl serum and 2 ml diethyl ether, and thereafter diluted 1:2 with the zero calibrator from the kit. In addition, two purified serum samples were serially diluted 1:2 with the zero calibrator down to the estimated detection limit.
Assay precision was assessed by use of pooled serum from children and one commercial quality control (Lyphochek from Bio-Rad). Each serum pool was measured as a duplicate in 10–20 different assays of at least four different kit lots. The interassay coefficient of variations (CV) were calculated from the mean of the duplicates. Clinical usefulness (usually termed functional sensitivity) was defined as the lowest concentration of an assay that can be measured with an interassay CV of 20%.
To assess accuracy, 17β-estradiol was removed by charcoal stripping in two different pooled child sera, originating from prepubertal children and children in early puberty. After charcoal stripping, the samples were divided in two, and 24 pmol/l 17β-estradiol was added (spiked) in one of each pair of charcoal stripped child sera. Charcoal stripped sera and spiked sera were thereafter measured with direct RIA and extraction RIA.
The concentration of sex hormone-binding globulin (SHBG) in the zero calibrator was determined in four different zero calibrator lots by RIA (Orion Diagnostica).
To test the clinical reliability of the extraction RIA, all 17β-estradiol results during the years 2004–2006 originated from girls with precocious puberty undergoing long-acting gonadotrophin-releasing hormone (GnRH) analog treatment. The goal of the treatment was to reversibly stop pubertal development by suppressing gonadotropins/17β-estradiol levels to the prepubertal range.
The AutoDELFIA assay The minimal detection concentration for the modified sensitive AutoDELFIA was calculated with the aid of a precision profile, defined as the lowest level that could be measured with an intra-assay CV of 10%. Assay precision was tested in four child serum pools: girls breast stages 1–2, girls breast stage 2, girls breast stages 2–4 (7), and one commercial control.
Comparison between extraction RIA and direct immunoassays
To evaluate the concordance between the direct immunoassays and the established extraction RIA, 275 serum samples were run in parallel with the direct RIA and the extraction RIA. The serum samples originated from 229 females aged 0–19 years (median 11 years) at pubertal stages 1–5 and 46 pooled child sera. In addition, 122 children's serum samples were run in parallel with the AutoDELFIA and the extraction RIA.
Statistical analysis
Linear regression analyses were performed on log-transformed data. Method comparison was evaluated by a Bland–Altman plot (10). Undetectable values, i.e., values below zero, were set to 1 pmol/l. A value of P<0.05 was considered significant (11).
| Results |
|---|
|
|
|---|
The modified Spectria's analytical detection limit was 9 pmol/l, calculated as the apparent concentration 3S.D. from the counts at maximum binding. The analytical detection limit was confirmed in dilution tests, in which purified, not concentrated, serum samples from children could be diluted down to 9 pmol/l and obtain expected 17β-estradiol concentrations (Fig. 1A). The purified samples showed linearity down to the detection limit of 9 pmol/l and even lower concentrations (Fig. 1B). An analytical detection limit of 9 pmol/l is equal to 4 pmol/l (1.2 pg/ml) for serum samples that have been concentrated in the diethyl ether extraction step prior to quantification by Spectria Estradiol RIA.
|
The results of imprecision studies conducted with child serum pools are summarized in Table 1. The serum pool originating from girls at breast stages 1–2, resulted in significantly higher 17β-estradiol concentrations when measured by the direct RIA and AutoDELFIA when compared with measurements by the extraction Spectria (n=20, P<0.001,+58% and n=5, P<0.001,+267% respectively).
|
Our laboratory participates in the Bio-Rad External Quality Assurance Services. For the year 2006, the direct 17β-estradiol RIA's and the extraction 17β-estradiol RIA's bias from the mean of 125 laboratory results were –24 and –26% respectively.
The SHBG concentrations in the zero calibrators range between 10 and 26 nmol/l (median 25 nmol/l).
Clinical usefulness
The functional sensitivity of the concentration where the interassay CV is 20% is estimated at 6 pmol/l (1.6 pg/ml) for the extraction RIA (Table 1). The 17β-estradiol concentrations quantified with the extraction RIA in samples from girls (aged 4–17 years) during prepuberty (with Tanner breast stage 1; n=47) and early puberty (Tanner breast stage 2; n=54) ranged between <4–25 and 5–88 pmol/l, with median values 6 and 16 pmol/l respectively. The corresponding values for the direct RIA was <9–46 and <9–70 pmol/l, with median values 12 and 14 respectively. We found a significant difference in 17β-estradiol concentrations between girls with breast stages 1 (prepubertal) and 2 (early puberty), when quantified by the extraction RIA (P<0.0001) but no difference when quantified with the direct RIA. Samples from girls with breast stage 1, measured by the direct RIA, were significantly higher than those measured by the extraction RIA (P<0.01).
The clinical reliability of the extraction RIA is demonstrated in Fig. 2. In total, 37 serum samples from girls with precocious puberty had been quantified for 17β-estradiol measurements at our laboratory. Seven of these samples were submitted due to clinical suspicion of lack of treatment efficacy (pubertal progress). All these samples resulted in 17β-estradiol concentrations in the pubertal range. Thirty samples were sent as a routine control. Out of 30 cases, 24 resulted in levels seen during prepuberty.
|
|
|
| Discussion |
|---|
|
|
|---|
Previously published papers have shown the importance of using an extraction step prior to immunoassay for determination of the low serum concentration present in postmenopausal women (3, 5). In our study, we show the importance of a preceding purification and concentration of the analyte prior to quantification by commercial RIA for determination of low serum 17β-estradiol concentrations in children. Together, these studies establish that direct immunoassays do not perform well at low 17β-estradiol concentrations, not even those with low analytical detection limit. This may be due to potentially interfering substances, particularly cross-reacting water-soluble steroid conjugates, binding proteins (especially SHBG), and non-specific binding (NSB). Concentrations of SHBG are known to be high in prepubertal children (median 92 nmol/l) (12, 13) and postmenopausal women (mean 103 nmol/l) (14), and in combination with the 5000–15 000 times lower 17β-estradiol concentrations it is a challenge to quantify accurate concentrations of 17β-estradiol. In this study, we show that the 17β-estradiol calibrators included in the Spectria kit on average contain four times lower SHBG concentrations when compared with that in children's sera. To obtain accurate results with a direct immunoassay, the concentration of SHBG, and NSB in a patient's serum samples ought to be similar to that in the calibrators (15, 16, 17). Using an extraction step prior to quantification by RIA, the patient's SHBG and NSB are replaced by a zero calibrator containing the same concentration of SHBG and NSB as is found in the calibrators in the kit. However, if the assay has problems with high cross-reactivity, an extraction does not overcome these problems. The Spectria assay that we use has very low cross-reactivity with estradiol-related steroids.
By charcoal stripping, it was confirmed that the interfering substances present in child sera are removed by the ether extraction procedure. Charcoal stripping removes 99% of initially present estradiol (18) and is a common procedure when producing zero calibrators. In this study, the expected 17β-estradiol concentration after charcoal stripping was 0 pmol/l and after spiking, the expected 17β-estradiol concentration was 24 pmol/l. The extraction RIA resulted in correct 17β-estradiol concentrations, both before and after spiking, but the direct RIA resulted in falsely high concentrations both after charcoal stripping and after addition of known amounts of 17β-estradiol.
The overestimation of 17β-estradiol concentrations in the direct immunoassays was also demonstrated by the impression data. The pooled serum samples originating from prepubertal girls and girls at breast stage 2 generally resulted in an overestimation. It is typical that the serum pools numbered 2 and 3 resulted in similar mean 17β-estradiol concentrations when measured by the direct RIA, but the serum pool originating from prepubertal children resulted in a much higher interassay CV than the serum pool originating from both prepubertal and pubertal girls. The results obtained by the AutoDELFIA are still overestimating and are similar to those obtained by the original method (9).
By directly or indirectly increasing the sample volumes and incubation times, we decreased the analytical detection limits from 30 and 50 pmol/l down to 9 and 11 pmol/l respectively, but we did not improve the clinical usefulness. The analytical sensitivity is estimated by determining the imprecision of the zero calibrator, which does not always correspond to the detection limit in the intended specimen. This is because the specimens are less pure and may contain interfering substances. In this study, purification by diethyl ether extraction of the children's serum samples was required to confirm the analytical sensitivity.
The most sensitive assays are considered to be the recombinant cell bioassays (RCBA). Klein and co-workers have developed a RCBA to measure estrogenic activity with a detection limit between 0.07 and 0.7 pmol/l 17β-estradiol equivalents. The sample volume used was 800 µl, the interassay CV was 13% at 7 pmol/l and increased to about 60% at 0.07 pmol/l (19, 20). Another RCBA has been developed by Paris et al. (21). The required sample volume is 800 µl and the detection limit is below 4 pmol/l. The interassay CV was 20% at 6 pmol/l. It has been shown previously that RCBA correlates to gas chromatography–tandem mass spectrometry (22). The extraction RIA corresponds well with RCBA for values down to the analytical limits (23). Furthermore, the 17β-estradiol values obtained in prepubertal girls with our extraction RIA are more or less identical to those obtained by RCBA (20, 21). By compilation of all this data, we estimate that our modified commercial extraction RIA has similar clinical usefulness as RCBAs.
There are three clinical situations in pediatrics where there is a need to determine low 17β-estradiol levels in girls to: 1) establish whether puberty has started, 2) monitor GnRH (lutenizing hormone releasing hormone) analog treatment of precocious puberty and 3) monitor hormone replacement therapy in hypogonadal children. The extraction RIA has the sensitivity to establish whether puberty has started or not and to determine the low 17β-estradiol levels seen at pubertal induction with transdermal estradiol patches in hypogonadal girls (6, 8). Figure 2 shows the results from 37 samples sent to our laboratory during the years 2004–2006 for therapy control of treatment of precocious puberty with long-acting GnRH analog. The goal of the treatment is to reversibly stop pubertal development, by suppressing gonadotropins/17β-estradiol levels to the prepubertal range (24). The state of the art method is to show that the efficacy of the treatment has been accomplished by GnRH challenge test. However, development of sensitive methods for determining 17β-estradiol makes it possible to assess the efficacy of the treatment more simply and inexpensively by determination of 17β-estradiol instead of gonadotropins after GnRH test (25). Some pediatricians have used this possibility and sent samples to our laboratory for analysis. In all seven cases where the reason for analysis was clinical suspicion of lack of efficacy, all the samples had pubertal levels. In 24 out of 30 cases where there was no information about the efficacy, serum 17β-estradiol concentrations were in the prepubertal range or under the detection limit of the assay. Even if the result is further confounded by not knowing the reasons for therapy control, the result lends further support to the theory that the assay works for use in a clinical setting. The extraction RIA is at present an accredited assay by SWEDAC in Sweden, SS-EN ISO 15189 (no. 1899).
In conclusion, commercially sensitive immunoassays do not give clinically useful results in sera from prepubertal children and children in early puberty. A purification step prior to commercially sensitive immunoassay is necessary to achieve clinically useful results when quantifying serum 17β-estradiol in children. Clinical laboratories should be aware that the clinical usefulness of their current procedures may be far from the method's analytical detection limit. It is important to validate each assay for its intended purpose.
| Acknowledgements |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. H. A. Ronkainen, V. Kovanen, M. Alen, E. Pollanen, E.-M. Palonen, C. Ankarberg-Lindgren, E. Hamalainen, U. Turpeinen, U. M. Kujala, J. Puolakka, et al. Postmenopausal hormone replacement therapy modifies skeletal muscle composition and function: a study with monozygotic twin pairs J Appl Physiol, July 1, 2009; 107(1): 25 - 33. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Ankarberg-Lindgren, E. Norjavaara, M. M. Kushnir, W. L. Roberts, A. L. Rockwood, J. Bergquist, A. M. Bunker, and A. W. Meikle Are Estradiol Results Determined by the Tandem Mass Spectrometry Assay Clinically Useful for Children? Am J Clin Pathol, May 1, 2009; 131(5): 746 - 750. [Full Text] [PDF] |
||||
![]() |
T. Finni, V. Kovanen, P. H. A. Ronkainen, E. Pollanen, G. R. Bashford, J. Kaprio, M. Alen, U. M. Kujala, and S. Sipila Combination of hormone replacement therapy and high physical activity is associated with differences in Achilles tendon size in monozygotic female twin pairs J Appl Physiol, April 1, 2009; 106(4): 1332 - 1337. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Allvin, C. Ankarberg-Lindgren, H. Fors, and J. Dahlgren Elevated Serum Levels of Estradiol, Dihydrotestosterone, and Inhibin B in Adult Males Born Small for Gestational Age J. Clin. Endocrinol. Metab., April 1, 2008; 93(4): 1464 - 1469. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||